PULSE 6 SERVICE MANUAL
PAGE 3.2
2009SUNRISE MEDICAL
Section 3
Check that the batteries are fully charged and in good
condition; and check all cables and connections. Check
the connections to the left motor, look for a loose or
damaged
connector.
Measure the resistance from the bottom contact of the
red thick wire on the 4-pin left motor connector to each
of the top contacts of the connector (figure A3.3.1).
Measure the resistance from the bottom contact of the
black thick wire on the 4-pin left motor connector to
each of the top contacts of the connector see (below
right). If all of the readings are open, then replace the
controller. If any of the readings are short, then replace
the left motor.