PULSE 6 SERVICE MANUAL
PAGE 5.7
2009SUNRISE MEDICAL
Tilt Will Not Operate
1. Selected tilt on the hand control (fi g. 5.9).
Operate the joystick and listen closely to the
control module for a click. This click indicates
that the control module is closing the power relay
inside and supplying power to the tilt actuator.
If no click is heard, replace the control module.
Refer to Control Module Replacement section this
manual. Retest as necessary. If click is heard
proceed to step 3.
2. Verify that tilt is selected on the hand
control. Operate the joystick and listen closely
to the tilt actuator. If the actuator appears to be
running but the tilt is not moving, replace the tilt
actuator. Refer to Tilt Actuator and Micro-Switch
Removal section of this manual. Retest as
necessary.
3. Locate the 6-pin connector on the rear of
the tilt (fi g. 5.10).
4. Disconnect the 6-pin cable leading to the
control module.
The tilt actuator used in the tilt system receives power through a 6-pin connector located at the rear
of the tilt. The actuator also contains micro-switches that open at the end of their stroke to prevent
stressing of the tilt system. Around these micro switches are diodes that allow reverse power to fl ow
to the actuator when reversing direction.
5. Select tilt on the hand control and place a
rubber band around the joystick to hold it displaced
(fi g. 5.11).
R-net Fault Codes cont.
Section 5
fig. 5.9
fig. 5.10
fig. 5.11