PULSE 6 SERVICE MANUAL
PAGE 5.5
SEPT 2011SUNRISE MEDICAL
3. Locate the creep micro-switch mounted at the
rear of the actuator "A" ( gure 5.5.1).
Note: The creep micro-switch may be seen from
the top of the seating system on the right rear of
the actuator. It may be necessary to remove the
seat pan to access the switch.
4. Manually operate the creep micro-switch while
observing the hand control or Omni for the turtle
symbol to go on and go off with the micro-switch
operation. If the symbol changes, the problem
is in the programming. Contact Sunrise Medical
Technical Service for assistance. If the symbol
does not change, proceed to step 5.
5. Lower the tilt system.
6. Locate the 6-pin connector "B" at the rear of
the tilt unit ( gure 5.5.2).
7. Disconnect the connector leading to the control
module.
8. Test the micro-switch wiring by performing
the following test.
a. Ensure that the seat is in the full down
position.
b. Use Multimeter and measure continuity
between the indicated pins below on the 6-pin
connector mounted on the tilt. When down the
indication should be open ( gure 5.5.3).
c. If it is closed, replace the micro-switch
assembly. Refer to Tilt Actuator and Micro-Switch
Removal section of this manual.
d. Retest as necessary.
9. If the above test passes, replace in this
order:
a. Cable leading from tilt to control module.
b. Control module
10. Retest as necessary
Open
R-net Fault Codes cont.
Section 5
fig. 5.5.3
fig. 5.5.2
fig. 5.5.1
A
B