PULSE 6 SERVICE MANUAL
PAGE 7.16
SEPT 2011 SUNRISE MEDICAL
Section 7
4. Reconnect the 6-pin connector on the rear of
the tilt and partially tilt the seating system back
to expose the actuator. Disconnect the 6-pin
connector
Note: Complete step 4 only if the micro-switch assembly is to be replaced separately.
6. Locate the micro-switch assembly at the rear
of the tilt (fi gure 7.16.2). Note the connections
leading to the micro-switch. Separate and label
the wires leading to the micro-switch. Reverse the
procedures to replace it and retest as necessary.
Warning: The tilt system is free to fall during
the next step. Secure the seating system prior
to removing the tilt actuator.
7. Use two 13 mm wrenches and remove the
nuts at both ends of the actuator (fi gure 7.16.3).
Remove both pins from the tilt actuator.
Removal Procedures (cont.)
fig. 7.16.2
fig. 7.16.1
fig. 7.16.3
5. Locate the tilt actuator under the seat
(fi gure 7.16.1).