PULSE 6 SERVICE MANUAL
PAGE 7.9
SEPT 2011SUNRISE MEDICAL
Section 7
Shock Removal
1. Remove the batteries. Refer to the Battery
Removal section.
2. Remove the motor/gearbox assembly from
the side on which you are removing the shock.
Refer to Motor/Gearbox Removal section of this
manual.
3. Use a 17 mm wrench and remove the top bolt
securing the shock to the base (fi gure 7.9.1).
4. Use two 19 mm wrenches and remove
the bottom nut and bolt securing the shock to the
base (fi gures . 7.9.2 & 7.9.3).
5. Lift the shock from the base.
6. Reverse the previous procedures to install the
shock.
Removal Procedures (cont.)
fig. 7.9.1
fig. 7.9.2
fig. 7.9.3
NOTE: The shocks used in this power base are
pre-adjusted at the factory. Contact Sunrise
Medical Technical Service if adjustment is
required.